We use cookies to personalise content and ads, to provide social media features and to analyse our traffic. We also share information about your use of our site with our social media, advertising and analytics partners who may combine it with other information that you’ve provided to them or that they’ve collected from your use of their services. You consent to our cookies if you continue to use our website.

Synchronizing Instrument Clocks: Details

This section describes how individual instruments’ clocks are set.
The delay introduced by LAN communications is not taken into account when setting instruments’ clocks.

ModelDetailed clock synchronization operation
PQ3100Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:[Ver. 2.10]
If the instrument is recording, recording is stopped when the clock is set and then restarted.
[Ver. 2.20 or later]
The clock is set without stopping recording, even if the instrument is recording.(*1) (*2)
PQ3198Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:The clock is set without stopping recording, even if the instrument is recording.(*1) (*3)
PW3360Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:[Ver. 3.20 or earlier]
If the instrument is recording, recording is stopped when the clock is set and then restarted.
[Ver. 3.21 or later]
The clock is set without stopping recording, even if the instrument is recording.(*1) (*4)
PW3365Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:[Ver. 2.00 or earlier]
If the instrument is recording, recording is stopped when the clock is set and then restarted.
[Ver. 2.10 or later]
The clock is set without stopping recording, even if the instrument is recording. (*1) (*4)
PW3390Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:00.000" (*5)
Operation during recording:If the instrument is integrating values, integration is stopped when the clock is set and then restarted.
PW6001, PW8001Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:If the instrument is integrating values, integration is stopped when the clock is set and then restarted.
LR8400,LR8401,LR8402,
LR8450, LR8450-01
LR8410, LR8416
Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:If the instrument is recording, recording is stopped when the clock is set and then restarted.
MR6000Referenct time:Computer time
Timing:When the reference time reaches "hh:mm:ss.000"
Operation during recording:If the instrument is recording, recording is stopped when the clock is set and then restarted.

(*1) If the reference time and instrument time differ by more than the recording interval, there may be a blank interval in the recording data. It is recommended to stop recording before synchronizing an instrument's clock.
(*2) Recording may stop and then restart when setting the clock in the following circumstances:

  • If the instrument is in the event IN, KEEP, or OUT state, or if the instrument is saving fluctuation data
  • If the interval save time is set to 150 or 180 cycles
  • If flicker is set to any value other than OFF
  • If the reference time and instrument time differ by more than 5 min.

(*3) Recording may stop and then restart when setting the clock in the following circumstances:

  • If the instrument is in the event IN, KEEP, or OUT state, or if the instrument is saving fluctuation data
  • If the reference time and instrument time differ by more than 1 min.

(*4) Recording may stop and then restart when setting the clock in the following circumstances:

  • If the reference time and instrument time differ by more than 5 min.

(*5) Setting the clock may result in a wait time of up to 1 min. as the system waits until XX:XX:00 to set the clock.