Models that support event recording and associated precautions
・When the "Maximum number of events that can be saved" has been recorded, the measurement will stop and no further events will be recorded.
・It may not be possible to detect events that last less than "Minimum detectable event duration" due to the inability to measure them accurately.
CM4001
・Supported functions: AC current measurement (ACA)
・Threshold setting: > (greater than)
・Maximum number of events that can be saved:99
・Minimum detectable event duration:200ms
CM4002,CM4003
・Supported functions: AC current measurement (ACA)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:999
・Minimum detectable event duration:200ms
FT6380-50
・Supported functions:AC current measurement(ACA)、Resistance measurement(Ω)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:25
・Minimum detectable event duration:1s
DT4261
・Supported functions:DC voltage measurement(DCV)、AC+DC voltage measurement(ACDCV)、DC voltage measurement(High V Probe Mode)(DCHV)、AC voltage measurement(ACV)、Frequency measurement(HzV)、Resistance measurement(Ω)、Electrostatic capacity measurement(F)、AC current measurement(with clamp sensor)(Clamp)、AC current measurement(ACA)、Frequency measurement(HzA)、DC current measurement(DCA)、AC+DC current measurement(ACDCA)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:99
・Minimum detectable event duration:200ms
CM4371-50,CM4373-50,CM4375-50
・Supported functions:DC voltage measurement(DCV)、DC voltage measurement(High V Probe Mode)(DCHV)、AC voltage measurement(ACV)、AC+DC voltage measurement(ACDCV)、Frequency measurement(HzV)、AC current measurement(ACA)、DC current measurement(DCA)、AC+DC current measurement(ACDCA)、Frequency measurement(HzA)、Resistance measurement(Ω)、Electrostatic capacity measurement(F)、Temperature measurement(℃/℉)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:99
・Minimum detectable event duration:400ms(Excluding some functions)※
※ Frequency、Electrostatic capacity:400ms ~ 4000ms(Depends on the measurement value)、Temperature:2000ms
CM4141-50
・Supported functions:DC voltage measurement(DCV)、DC voltage measurement(High V Probe Mode)(DCHV)、AC voltage measurement(ACV)、AC+DC voltage measurement(ACDCV)、Frequency measurement(HzV)、AC current measurement(ACA)、Frequency measurement(HzA)、Resistance measurement(Ω)、Electrostatic capacity measurement(F)、Temperature measurement(℃/℉)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:99
・Minimum detectable event duration:400ms(Excluding some functions)※
※Frequency、Electrostatic capacity:400ms ~ 4000ms(Depends on the measurement value)、Temperature:2000ms
CM3286-50
・Supported functions:AC current measurement(ACA)、AC voltage measurement(ACV)
・Threshold setting: > (greater than)、<(less than)
・Maximum number of events that can be saved:99
・Minimum detectable event duration:1s